MedRx Hearing Aid Analyzer
MedRx Hearing Aid Analyzer
넳 넲

MedRx Hearing Aid Analyzer

The historical MedRx page describes AVANT HIT+ as a compact hearing-aid analyzer for on-site performance checks. It says the measurements can help investigate reported background noise, battery use, sound distortion or feedback; these are intended uses, not a promise to resolve a user's complaint.

MedRx Hearing Aid Analyzer

Model: AVANTHIT+

Signal types
Pure tone; wideband noise including pink, white, Byrne LTASS and ANSI-weighted noise; chirp; ICRA
Hearing-aid analysis standards printed
ANSI S3.22-2009; IEC 60118-7:200
Loudest output as printed
≥95 dB SPL
Analog battery output
0.1–1.5 V adjustable in 0.1 V steps
Current test range
20 µA–20 mA; ±5% accuracy
Analog battery models
A10; A13; A312; A675
Frequency range
125–8,000 Hz
Test microphone
Maximum output sound pressure level 140 dB SPL; 4 × 20 dB attenuation; 2 cc coupler
Reference microphone
Coupler and other adapters listed
Magnetic induction coil
Maximum sound output 100 dB SPL; omnidirectional microphone; 31.6 mA induction coil
LED indicator
Green light signifies availability
Communication port
USB 2.0, compatible with USB 1.1
Power supply
USB and 5 V DC; printed current below 40 mA
Dimensions
Approximately 24 × 25 × 16 cm
Standard configuration
AVANT HIT+ main unit; test microphone; reference microphone; 2 cc coupler and accessories; four analog battery sets; USB data cable
Home    Products    MedRx Hearing Aid Analyzer

The historical MedRx page describes AVANT HIT+ as a compact hearing-aid analyzer for on-site performance checks.