MedRx Hearing Aid Analyzer
The historical MedRx page describes AVANT HIT+ as a compact hearing-aid analyzer for on-site performance checks. It says the measurements can help investigate reported background noise, battery use, sound distortion or feedback; these are intended uses, not a promise to resolve a user's complaint.
MedRx Hearing Aid Analyzer
Model: AVANTHIT+
- Signal types
- Pure tone; wideband noise including pink, white, Byrne LTASS and ANSI-weighted noise; chirp; ICRA
- Hearing-aid analysis standards printed
- ANSI S3.22-2009; IEC 60118-7:200
- Loudest output as printed
- ≥95 dB SPL
- Analog battery output
- 0.1–1.5 V adjustable in 0.1 V steps
- Current test range
- 20 µA–20 mA; ±5% accuracy
- Analog battery models
- A10; A13; A312; A675
- Frequency range
- 125–8,000 Hz
- Test microphone
- Maximum output sound pressure level 140 dB SPL; 4 × 20 dB attenuation; 2 cc coupler
- Reference microphone
- Coupler and other adapters listed
- Magnetic induction coil
- Maximum sound output 100 dB SPL; omnidirectional microphone; 31.6 mA induction coil
- LED indicator
- Green light signifies availability
- Communication port
- USB 2.0, compatible with USB 1.1
- Power supply
- USB and 5 V DC; printed current below 40 mA
- Dimensions
- Approximately 24 × 25 × 16 cm
- Standard configuration
- AVANT HIT+ main unit; test microphone; reference microphone; 2 cc coupler and accessories; four analog battery sets; USB data cable
The historical MedRx page describes AVANT HIT+ as a compact hearing-aid analyzer for on-site performance checks.